Meeting Banner
Abstract #1593

A Wireless Receive Surface Coil at 1.5T

Busra Kahraman-Agir1, Rawish Roshan Bansropansingh1, Mark Gosselink1, and Dennis Klomp1
1Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands

Synopsis

Keywords: Non-Array RF Coils, Antennas & Waveguides, Non-Array RF Coils, Antennas & Waveguides, wireless coils, inductive coupling

Motivation: MRI is a highly preferable but also an expensive modality mainly due to 1)high operational cost, 2)staff cost, 3)long scan and preparation times.

Goal(s): Reducing costs and simplifying MRI scans.

Approach: A wired coil and a wireless coil which was inductively coupled to a receiver loop were compared in terms of their efficiency and SNR results.

Results: On the bench, a 0.8 dB+/-0.5dB of decrease was observed in the efficiency of the wireless coil over a range of 3 cm. The SNR of the wireless coil was practically identical to the wired coil both in phantom and in vivo MR images.

Impact: The efficiency obtained by a wired coil decreased by 0.8 dB when an inductively coupled wireless coil was used. Phantom and in vivo measurements show a practically identical sensitivity profile and SNR for both coils.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords