A Novel Transceive 2-channel Orthogonally Arranged Dual Loop Solenoid RF Coil Design for Improving Low Field MRI SNR

Yan Gu1, Hong Hua Pei2, Xin Qi2, Ren Jie Zong2, Jian Bao2, Shao Ying Huang3, and Bing Keong Li2
1The First People’s Hospital of Lianyungang, Jiangsu Province, Lianyungang, China, 2Jiangsu LiCi Medical Device Co. Ltd, Lianyungang, China, 3Singapore University of Technology and Design, Singapore, Singapore

Synopsis

Keywords: Low-Field MRI, Low-Field MRI

Motivation: There are growing interest in low field specialised MRI R&D but they are currently burdened by their low SNR images.

Goal(s): A novel 2-channel orthogonally arranged dual loop solenoid RF coil design has been proposed to improve low field MR image SNR.

Approach: Two transceive 2-Channel RF coil prototypes using the novel design and using conventional design were constructed and tested in a 0.35T MRI system.

Results: In comparing the acquired MR images, it is shown that the proposed novel design can improve low field MR image SNR by ≥20% as compared to the conventional design.

Impact: A novel 2-channel orthogonally arranged dual loop solenoid transceive RF coil has been constructed and tested in a 0.35T MRI system. Using this new RF coil design it is shown that the acquired image SNR can be increased by ≥20%.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords