Effect of LCModel basis set B0 frequency on MRS quantification

Dinesh K Deelchand1 and Gülin Öz1
1Radiology, University of Minnesota, Minneapolis, MN, United States

Synopsis

Keywords: Spectroscopy, Spectroscopy, LCModel

Motivation: The actual B0 resonance field of 3T scanners is different between the three major MR vendors.

Goal(s): Here, we report the effect on metabolite quantification of using basis sets at different B0 fields than the B0 frequency of the measured in vivo MRS data.

Approach: Basis sets were simulated at thirteen different B0 fields. Semi-LASER MRS data measured from the posterior cingulate cortex at 3T were fitted using LCModel.

Results: Results show that biases in metabolite concentrations were within 1% when the basis set B0 frequencies were within ±0.5 MHz from the actual scanner frequency.

Impact: A single sequence-specific basis set can be used to analyze harmonized spectroscopic data collected on clinical 3T scanners from different vendors that operate at different frequencies.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords