A fully automated implant mode MRI scan workflow for asymptomatic patients.

Kavitha Manickam1, Chitresh Bhushan2, Dawei Gui1, Maggie Fung3, Shiv Kaushik1, Eric Fiveland2, Dattesh D Shanbhag4, and Hollis Potter5
1GE HealthCare, Waukesha, WI, United States, 2GE HealthCare, Niskayuna, NY, United States, 3GE HealthCare, New York, NY, United States, 4GE HealthCare, Bangalore, India, 5Hospital for Special Surgery, New York, NY, United States

Synopsis

Keywords: Other Musculoskeletal, MSK, Implant mode scan, Prescan, Calibration, Mavric

Motivation: Patients are often unaware of the presence and location of metal implants, which may be confirmed only after localizer scan, thus potentially disrupting workflow.

Goal(s): To detect the presence of metal implant within the first few seconds of prescan and further characterize the location and shape of the metal from the calibration.

Approach: Prescan data of phantom with metal screws and human volunteer with metallic implants were used.

Results: The skewness of the center frequency signal is from 1.37 to 4.47 for implants, while for normal subjects, the signal is very symmetric and the calculated skewness is <0.3.

Impact: Automated MR scanning workflow for the patients with implants to improve efficiency of the MR technologists.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords