Meeting Banner
Abstract #2829

A comparison of coil loading and SNR for elliptical solenoidal coil vs. a tight fitting dome helmet coil for low field neuroimaging

Tom O'Reilly1 and Andrew Webb1
1Leiden University Medical Center, Leiden, Netherlands

Synopsis

Keywords: Low-Field MRI, Low-Field MRI

Motivation: The narrow bandwidth of RF coils for low field MRI means that coils often require re-tuning between subjects due to different loading conditions.

Goal(s): We compare the performance and loading effects of an elliptical solenoid and a dome helmet coil to see which coil is more sensitive to loading

Approach: The SNR and transmit efficiency of the two coil types are compared in when lightly loaded (phantom) and normally loaded (in-vivo) settings

Results: The elliptical solenoid coil has a substantially smaller frequency shift when loaded (~1kHz vs ~13kHz) and has superior SNR in-vivo compared to a tight -fitting dome helmet coil.

Impact: We compare the performance of tight-fitting dome helmet coil to an elliptical solenoid coil and find that due to reduced loading the performance of the elliptical coil is superior in-vivo while having better frequency stability with differing loads.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords