Keywords: Data Acquisition, Data Acquisition, Noise, SNR, detectability, estimation
Motivation: Pixel SNR is a common metric for sequence design and scan implementation, however it is not ideal for detectability and estimation.
Goal(s): To take and modify an existing noise measure from the literature and test it against pixel SNR for both white and red (high spatial frequency) noise.
Approach: Ten individuals performed a detection task on 100 synthetic low SNR images with white or red noise.
Results: The new metric was more predictive of detection than pixel SNR or circle radius. Images with red noise had worse pixel SNR but better detection rates.
Impact: A proposed SNR metric is more relevant than pixel SNR for both detection and estimation, and consistent with our findings of better detectability in red noise over white noise. This is important for the design of clinical pulse sequences.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.