Keywords: Data Acquisition, Gradients, Gradient delay estimate
Motivation: The study was motivated by substantially reducing the calibration scan time for gradient delay estimation in spiral imaging.
Goal(s): we only acquire a single spiral readout along each channel for delay estimation, which enables a substantial reduction in the calibration scan time.
Approach: Our proposed method of SODA estimates the delay by calculating the relative shift between the MR spectrums required in a single scan as the k-space trajectory progresses from the negative kx direction to the positive kx direction and vice versa, i.e., from +kx to -kx.
Results: The delay values demonstrated robust agreement and consistency between SODA and the reference.
Impact: The SODA method delivers accurate delay estimates within extremely short scan time (about 50 ms). This technique can seamlessly integrate with any host sequence to correct k-space location errors stemming from gradient delays.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.