Keywords: Pulse Sequence Design, Pulse Sequence Design
Motivation: Robust measures of the relaxation rate R2* are essential for use in neuroscience studies. With standard acquisition techniques, data are sampled consecutively at multiple echo times at the same phase of the cardiac cycle. This coherent sampling leads to a high level of cardiac-induced noise in brain maps of R2*.
Goal(s): We design a new data acquisition strategy that mitigates the impact of cardiac-induced noise in brain quantitative R2* maps.
Approach: We incoherently sample cardiac-induced noise by shifting k-space location during the multi-echo acquisition.
Results: Compared to standard techniques, this strategy reduces the variability of R2* estimates across repetitions by 30-40%.
Impact: The proposed k-space shifting acquisition strategy reduces the level of cardiac-induced noise in brain maps of R2*. This will increase the sensitivity of brain change analyses in future neuroscience studies that include R2* mapping.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.