Artery selective 3D TOF by Using Asymmetrically RF-shimmed Pre-saturation pulse with 4-channel RF Transmit at 3T

Kosuke Ito1, Atsushi Kuratani1, Yukio Kaneko2, and Masahiro Takizawa1

1Healthcare Company, Hitachi Ltd., Chiba, Japan, 2Research and Development Group, Hitachi Ltd., Tokyo, Japan

A technique of ICA-selective 3D TOF imaging using asymmetrically RF-shimmed Pre-saturation pulse is presented. By using 4-channel RF transmit coil, spatially asymmetric RF transmission is achieved. Then apply RF shimming parameter to Pre-saturation pulse, the saturation effect becomes spatially asymmetric, and achieved selectively visualize blood flow from each ICA. This technique provides high contrast between blood and brain parenchyma because it does not make TR longer.