Meeting Banner
Abstract #0216

Concurrent Higher-Order Field Monitoring for Routine Head MRI: An Integrated Heteronuclear Setup

Christoph Barmet1, Bertram Jakob Wilm1, Matteo Pavan1, Georgios Katsikatsos1, Jochen Keupp2, Giel Mens3, Klaas Paul Pruessmann1

1Institute for Biomedical Engineering, ETH and University, Zurich, Switzerland; 2Philips Research Europe, Hamburg, Germany; 3Philips Healthcare, Best, Netherlands


A higher-order concurrent field monitoring setup is introduced for routine head MRI. It enables the tracking of dynamic field evolution up to 3rd order concurrently with data acquisition. This is particularly important for non-reproducible field contributions, e.g. due to magnet heating, breathing or external fields. The field information allows for the correction of image artifacts at the reconstruction stage.

Keywords