short VAPOR-like water suppression sequence is presented, exhibiting
similarly low B1 and T1 sensitivity and improved excitation profiles. The
improvements are based on optimization of flip angles and pulse durations of
chemical-shift selective pulses interleaved with fixed short delays. The
sequence consists of 6-pulse water presaturation with asymmetric RF pulses,
followed by B1-insensitive inversion and the localization module. The
improved robustness may be utilized for an accurate control of residual water
signal and exploiting it as a reference. Thanks to the reduced length and
reduced impact on metabolites, the sequence should improve quantifiability
and be suitable for spectroscopic imaging.