Andreas
Greiser1, Mehmet Akif Gulsun2, Arne Littmann1,
Jens Guehring2, Edgar Mueller1
1Siemens AG Healthcare Sector,
Vector-encoded
MR phase contrast acquisitions covering a bigger volume show larger errors in
velocity due to eddy currents and gradient non-linearities. A new scan method
is presented that acquires the volumetric dataset in multiple z-isocentered
steps. The resulting corrected velocity images and the influence on the flow
quantification results of the descending aorta and flow field visualization
were analyzed. The multiply isocentered approach results in an overall
increase of peak velocity estimates and flow values. The pixelwise standard
deviation of the calculated background
phase correction matrices across slices for fixed table position was 1.76
cm/sec vs. 0.61 cm/sec for z-isocentered.