Daniel Wilson1, J. Ashley Guthrie2,
Janice Ward2
1Medical Physics, Leeds Teaching
Hospitals NHS Trust, Leeds, West Yorkshire, United Kingdom; 2Radiology,
Leeds Teaching Hospitals NHS Trust, Leeds, West Yorkshire, United Kingdom
The
liver ADC was measured in 3 different positions in the liver in 33 subjects.
The mean ADC was signficantly different between the 3 different positions.
This is attributed to the presence of noise in the middle of the liver and
artefactual signal loss at high b-values at superior positions in the left
lobe. Care must be taken when measuring ADC in the liver.
Keywords