Fast T1 Mapping at 7T Using Look-Locker TFEPI

Emma Louise Hall1, Ali M. Al-Radaideh1, Su Y. Lim2, Susan T. Francis1, Penny A. Gowland1

1Sir Peter Mansfield Magnetic Resonance Centre, University of Nottingham, Nottingham, Nottinghamshire, United Kingdom; 2Clinical Neurology, University of Nottingham, Nottingham, Nottinghamshire, United Kingdom


Ultra high field has the benefit of increased SNR to facilitate high resolution imaging. However, the lengthened relaxation time requires long scan times to produce high resolution T1 maps due to the need to allow the system to return to equilibrium. Here we present a Look-Locker TFEPI sequence that allows the acquisition of high resolution, 1.25mm isotropic, T1 maps with large volume coverage at 7T in less than 6 minutes.