Jochen Mosig1,
  Achim Bahr1, Thomas Bolz1, Andreas Bitz2,
  Stephan Orzada2
1RF&Dosimetry, IMST GmbH,
  Kamp-Lintfort, Nordrhein-Westfalen, Germany; 2Erwin L. Hahn
  Institute for Magnetic Resonance Imaging, Essen, Nordrhein-Westfalen, Germany
In
  a previous work the design and numerical results for a composite right/left-handed
  metamaterial coil element were presented. This work shows dosimetric
  measurement and first imaging results together with further numerical
  results. A good agreement between the simulations and the measurements was
  observed. In contrast to the homogeneous B1 field, the circular polarized B1+
  field shows some discontinuities. This has lead to the investigation of
  different designs for the metamaterial element. As a result, an extended
  layout is presented, that eliminates the local minima in the field
  distribution of the original element, and shows a significant different field
  distribution.
Keywords