Meeting Banner
Abstract #2003

Testing the Variability of Diffusion Spectrum Imaging (DSI): Inter- & Intra-Site Comparison on Identical 3T Scanners

Alia Lemkaddem1, Alessandro Daducci1, Serge Vullimoz2, Margitta Seeck2, Francois Lazeyras3, Reto Meuli4, Gunnar Krueger5, Jean-Philippe Thiran1

1Signal Processing Laboratories (LTS5), Ecole Polythechnique Fdral de Lausanne (EPFL), Lausanne, Switzerland; 2Presurgical Epilepsy Evaluation Unit, Hpitaux Universitaires et facult de mdecine de Genve (HUG), Switzerland; 3Department of Radiology, Hpitaux Universitaires et facult de mdecine de Genve (HUG), Switzerland; 4Department of Radiology, University Hospital Center & University of Lausanne (CHUV), Switzerland; 5Advanced Clinical Imaging Technology, Siemens Medical Solutions-CIBM, Switzerland


The purpose of this study was to evaluate if it was possible to merge DSI data from two different sites. Therefore, we assessed inter- and intra-site reproducibility of two identical 3T scanners at two different hospitals in six healthy subjects. For investigating the reproducibility we used FA, ADC and GFA as measurements of the diffusivity. To validate our results we decided to use three different methodologies: A Tract-based, Regional based and whole brain voxel-based analysis. Our findings support the feasibility of cross-site pooling of DSI data from identical scanners.

Keywords