Novel FSE Method to Improve T2map Accuracy by Calculating B1map Simultaneously

Yoshihiro Tomoda1, Hitoshi Ikeda1

1MR Engineering, GE Healthcare Japan, Hino, Tokyo, Japan


Novel FSE method is proposed to improve the T2map accuracy by calculating B1map simultaneously without scan time elongation. Compared with the legacy CPMG method, T2 estimation error is low enough in a wider range of T2 and refocus FA. In addition, lower refocus FA is applicable to the proposed method, which would decrease SAR and increase the number of max slices per acquisition.