Abstract #0599
Which parameters are optimal? - A comprehensive numerical analysis of background phase correction with SHARP
Ferdinand Schweser 1 , PĪ¹nar Senay zbay 2,3 , Andreas Deistung 1 , Edsel Daniel Peres Gomez 1 , Xiang Feng 1 , Daniel Nanz 2 , and Jrgen R Reichenbach 1
1
Medical Physics Group, Institute of
Diagnostic and Interventional Radiology I, Jena
University Hospital - Friedrich Schiller University
Jena, Jena, Germany,
2
Department
of Radiology, University Hospital Zurich, Zurich,
Switzerland,
3
Institute
for Biomedical Engineering, University and ETH Zurich,
Zurich, Switzerland
Sophisticated harmonic artifact reduction for phase data
(SHARP) is regarded as one of the most important
techniques in the area of background correction for QSM.
The influence of the SHARP parameters has not been
investigated in detail so far and they are usually
chosen empirically, which may have serious consequences
for the comparability of studies relying on SHARP
processed phase images. We performed a comprehensive
analysis of the impact of different parameter choices on
the resulting images and propose a universal definition
of the regularization parameter that does not depend on
the spherical radius or on the image resolution.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.