B 0 Drift and Respiratory Motion Correction by Dual-Echo Susceptibility Correction (DESC)

Di Xu 1,2 , Steven M Shea 1,3 , Wesley D Gilson 1 , and Sunil G Patil 1

1 Center for Applied Medical Imaging, Imaging and Computer Vision, Siemens Corporation, Corporate Technology, Baltimore, Maryland, United States, 2 Department of Biomedical Engineering, Johns Hopkins University, School of Medicine, Baltimore, Maryland, United States, 3 Department of Radiology, Stritch School of Medicine, Loyola University Chicago, Maywood, IL, United States

A new B 0 correction method known as Dual-Echo Susceptibility Correction (DESC) for Echo-planar (EP) based magnetic resonance imaging (MRI) is introduced. The proposed DESC method is comprised of 1) an acquisition of two EPI echoes; 2) modeling the susceptibility changes from B0 drift and respiratory motion using a smoothing polynomial function and 3) an iterative reconstruction framework to accurately estimate B0 corrected data. The DESC method is evaluated in an MR thermometry application as a proof of concept. It is shown that the DESC method reduces B 0 drift and respiratory motion-related inaccuracies, while being computationally efficient.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords