Abstract #1402
Evaluation of MR Compatibility of a SiPM-based PET scanner for simultaneous PET/MR studies operating at animal 7-T MR scanner
Guen Bae Ko 1 , Daehong Kim 2 , Hyun Suk Yoon 1 , Min Sun Lee 1 , In Chan Song 3 , and Jae Sung Lee 1
1
Department of Nuclear Medicine, Seoul
National University, Seoul, Korea,
2
Molecular
Imaging and Therapy Branch, National Cancer Center,
Goyang, Gyeonggi, Korea,
3
Department of
Radiology, Seoul National University, Seoul, Korea
Here, wed like to report relevant influences of novel
SiPM-based PET scanner on MR images at ultra-high
magnetic field. Several MR image including FSE, SE, GRE,
3D SPGR, and EPI were evaluated with a 7-T MRI scanner
and 35-mm-inner-diameter mouse body transceiver coil.
The experimental results show that the PET insert unit
affects only minor effect on MR images that can be
ensured for preclinical evaluation. It was also proved
that to perform multi-functional study of PET and
functional MRI using our PET insert unit is possible.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Click here for more information on becoming a member.