Meeting Banner
Abstract #1461

Correction of 2D RF Pulses

Yuval Zur 1

1 GE Healthcare, Tirat Carmel, Israel

Two dimensional (2D) RF pulses with EPI excitation trajectory are extremely sensitive to system imperfections such as eddy currents and waveform distortions. These imperfections cause stop band excitation and pass band saturation. A method to correct these 2D RF pulses is presented. The correction is done by adding a phase to even sub pulses and gradient blips to the oscillatory excitation gradient. The added phase and the area of the blips are determined by a calibration done at system installation. The method was applied successfully to spectral spatial RF pulses at oblique and non oblique slice orientations.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords