Artifact correction based on diffusion coefficient

Gregory Lemberskiy 1,2 , Dmitry S Novikov 1 , and Els Fieremans 1

1 New York University School of Medicine, Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York, NY, United States, 2 Sackler Institute of Graduate Biomedical Sciences, New York University School of Medicine, New York, NY, United States

DWI images in the body are often afflicted with systematic vibration artifacts (SVA), characterized by localized regions of signal dropout. Generalized image processing techniques, such as outlier detection and histogram-based methods, require either large datasets or cannot account for nonlinear phenomena such as RF inhomogeneity. To overcome these limitations, we propose a rejection scheme based on unphysical diffusion coefficient values of the directional ADC, called Artifact Correction based on Diffusion Coefficient (ACDC). ACDC does not require many directions or averages to be efficient. Hence, its implementation in clinical systems would be straightforward.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords