A 10-channel TMS-compatible planar RF coil array for human brain MRI at 3T

Pu-Yeh Wu 1 , Ying-Hua Chu 1 , Aapo Nummenmaa 2 , Thomas Witzel 2 , Shang-Yueh Tsai 3 , Wen-Jui Kuo 4 , and Fa-Hsuan Lin 1,2

1 Institute of Biomedical Engineering, National Taiwan University, Taipei, Taiwan, 2 Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3 Institute of Applied Physics, National Chengchi University, Taipei, Taiwan, 4 Institute of Neuroscience, National Yang Ming University, Taipei, Taiwan

A slim TMS-compatible MRI 10-channel RF coil array was developed. The coil array was 2 mm thick to maximize TMS efficacy and positioning freedom. Empirical data show that the average SNR of TMS-compatible coil array is 44% better than the commercial 32-channel array within the 5 cm depth. This TMS-compatible RF coil array is expected to offer both anatomical images and functional MRI signal induced by TMS with high sensitivity.