Comparison of the accuracy in 2D and 4D PCMRI to evaluate oscillating flow in small diameters

Gwenael Page 1 , Roger Bouzerar 1 , Dominique Haye 2 , Dong-Joo Kim 3 , Hack-Jin Lee 3 , Anne-Virginie Salsac 4 , and Olivier Baledent 1

1 BioFlow Image, CHU Amiens, Amiens, France, 2 PFT Innovaltech, France, 3 Department of Brain and Cognitive Engineering, Korea, 4 Laboratoire de Biomecanique et Bioengenierie, CNRS, France

The aim of this study is to compare a high spatial resolution 2D PC-MRI protocol with a 4D PC-MRI protocol in a phantom with calibrated thin tubes and physiological oscillating flows. Protocols were applied on the phantom in a 3T, Achieva dStream, Philips using 32 head coils channels. Accuracy of the flow measurement was then evaluated by calculating the difference between expected and measured flow in each branches. For both sequences we had an error less than 10%, however 2D PC-MRI measurement was more accurate. It is possible to have 4D PC-MRI flow measurement consistent with 2D PC-MRI measurement