Constrained optimization of gradient waveforms for isotropic diffusion encoding

Jens Sjlund 1,2 , Markus Nilsson 3 , Daniel Topgaard 3 , Carl-Fredrik Westin 1,4 , and Hans Knutsson 1,5

1 Linkping University, Linkping, Sweden, 2 Elekta Instrument AB, Stockholm, Sweden, 3 Lund University, Sweden, 4 Brigham and Womens Hospital and Harvard Medical School, MA, United States, 5 Center for Medical Image Science and Visualization (CMIV), Linkping, Sweden

We propose a new framework for numerical optimization of diffusion encoding gradient waveforms. The formulation allows explicit control of hardware requirements, including maximum gradient amplitude, slew rate, heating and positioning of RF pulses. The power of the approach is demonstrated by a comparison with previous work on optimization of isotropic diffusion sequences, showing possible gains in diffusion weighting or in heat dissipation.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords