Abstract #3713

High resolution T1 mapping within seconds: model-based reconstruction without regularization

Volkert Roeloffs 1 , Xiaoqing Wang 1 , Tilman Sumpf 1 , and Jens Frahm 1

1 Biomedizinische NMR Forschungs GmbH, Max Planck Institute for Biophysical Chemistry, Gttingen, Niedersachsen, Germany

In this work we present a model-based reconstruction algorithm for the reconstruction of high resolution T1 maps from a fast spoiler-gradient free radial IR-FLASH sequence. By using minimal TR and formulating the problem in the parameter domain, no further regularization techniques are necessary. The resulting non-linear optimization problem is solved by the Gauss-Newton method. A phantom study revealed good agreement of the T1 values with the gold standard and a first in vivo experiment using the same parameter set as in the phantom study proofed the proposed technique to be robust.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords