Meeting Banner
Abstract #1530

Simultaneous Multi-Angular Relaxometry of Tissue with Magnetic Resonance Imaging (SMART MRI)

Alexander L Sukstanskii1, Jie Wen1, Anne H Cross2, and Dmitriy A Yablonskiy1

1Mallinckrodt Institute of Radiology, Washington University, Saint Louis, MO, United States, 2Neurology, Washington University, Saint Louis, MO, United States

The cross-relaxation effects between “free” and “bound” water affect the gradient recalled echo (GRE) MRI signal and can bias quantitative measurements of tissue relaxation parameters. Herein we have generalized the classical Ernst equation for the GRE signal dependence on sequence parameters (echo and repetition times, flip angle) by accounting for cross-relaxation effects. The derived equation creates a basis for a new technique - Simultaneous Multi-Angular Relaxometry of Tissue with Magnetic Resonance Imaging (SMART MRI). The technique allows simultaneous quantitative measurements of longitudinal and transverse relaxation rates constants and some essential cross-relaxation parameters without utilizing off-resonance magnetization transfer pulses.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords