Meeting Banner
Abstract #2176

Analysis of Eddy Currents for High Field RF Coil Design

Yu Li1, Fangfang Tang1, Bassem Henin1, Fabio Freschi1,2, Feng Liu1, and Stuart Crozier1

1School of ITEE, The University of Queensland, Brisbane, Australia, 2Department of Energy, Politecnico di Torino, Torino, Italy

Eddy currents are inevitably induced in the electrically conductive surroundings including the magnet cryostat vessel, RF coil, RF shield and other peripheral metallic structures. This results in image distortion and artefacts. In order to control eddy currents, some studies have also discussed the issue of eddy currents on magnet cryostat vessel and RF shields. In this work, the eddy currents on a 12 channel micro-strip RF array head coil for 7T MRI were analysed and compared with that of a slotted, double-sided copper RF shield. A preliminary approach is proposed to reduce the eddy current effect, without compromising RF performance.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords