Patient Factors and Electrode Placement Factors that Cause 3T-CMR to Poorly Synchronize with Waveforms of Vectorcardiography – Examination of Factors Worsening R/T-wave Amplitude Ratio –

KEI FUKUZAWA1, Rieko Ishimura2, Takashi Yoshida1, Chiharu Yoshihara1, Junji Takahashi1, and Masakatsu Tano1

1Radiology, Toranomon Hospital, Tokyo, Japan, 2Cardiovascular center, Toranomon Hospital, Tokyo, Japan

The lower R/T ratio inside a bore are obtained in patients with bathycardia, a condition in which the heart axis is close to the body axis. In addition, the short distance between the apex of the heart and the apex electrode and closeness of lead II and the heart axis are factors of a worse R/T ratio. These two points should be improved to obtain waveforms with a high R/T ratio.