Precision and Accuracy of Multi-Echo Based T2 mapping: Fitted vs Measured B1

Christopher L Lankford1 and Mark D Does1,2

1Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 2Electrical Engineering, Vanderbilt University, Nashville, TN, United States

This abstract provides a simple criterion by which to determine whether refocusing pulse flip angle should be fitted or constrained to a measured B1+ map in multiple spin echo (MSE) based T2 mapping protocols. Using propagation of error theory including the Cramér-Rao lower bound, it was discovered that T2 estimate precision is improved through constraint when the signal-to-noise ratio (SNR) of the B1+ map is at least one-half the SNR of the MSE measurement. The accuracy cost of constraint is also evaluated and presented.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords