Prior Image based Temporally Constrained Reconstruction for Magnetic Resonance guided HIFU

Jaya Prakash1, Nick Todd2, and Phaneendra K Yalavarthy3

1Institute for Biological and Medical Imaging, Helmholtz Zentrum Munich, Munich, Germany, 2Wellcome Trust Centre for Neuroimaging, UCL Institute of Neurology, University College London, London, United Kingdom, 3Supercomputer Education and Research Centre, Indian Institute of Science, Bangalore, India

Prior image based temporally constrained reconstruction (PITCR) algorithm was found to obtain accurate temperature maps with better volume coverage, and spatial, and temporal resolution than other algorithms using highly undersampled data in magnetic resonance (MR) thermometry. PITCR method is compared with the standard temporally constrained reconstruction (TCR) algorithm using ex-vivo pork muscle sample. It was seen that the PITCR method showed superior performance compared to the TCR approach with highly undersampled datasets.