Quantitative Susceptibility Mapping – Comparison of Silent and Conventional Acquisitions

Pauline W Worters1, Dirk Beque2, Robert D Peters3, Dominic Graziani4, Michael Carl5, Graeme C McKinnon3, and Christopher J Hardy4

1GE Healthcare, Menlo Park, CA, United States, 2GE Global Research, Munich, Germany, 3GE Healthcare, Waukesha, WI, United States, 4GE Global Research, Niskayuna, NY, United States, 5GE Healthcare, San Diego, CA, United States

A silent 3D multi-echo gradient-echo pulse sequence is developed for quantitative susceptibility mapping (QSM). The new, silent sequence is compared to the standard acquisition and gives comparable susceptibility values. The silent acquisition gives the benefit of patient comfort and workflow ease at some cost to SNR and acquisition time.