Meeting Banner
Abstract #1226

Universal Size-Optimized 48-Channel Phased-Array Receive Head Coil for 3T Clinical fMRI/Silent Imaging Application

Yun-Jeong Stickle1, Victor Taracila1, Sarah Tenley1, Clyve Follante1, Balint Franko1, Nabeel Malik1, Patrick Quarterman2, Fotis Vlachos2, and Peter Roemer2

1Engineering, GE Healthcare Coils, Aurora, OH, United States, 2Engineering, GE Healthcare, Waukesha, WI, United States

Typical high density phased-array head coils are design to fit tightly around the head to achieve increasing SNR and improving acceleration limiting space and excluding its use for larger patients. This study shows the results for an optimized 48-Channel phased-array receive head coil design on a universally sized football helmet-shaped former (fitting to more than 99th - percentile US male) providing similar SNR and improved acceleration compared to a 32-Channel close-fitting design coil. The performance of this array is demonstrated in highly accelerated head images on variously sized human subjects. This coil is optimized for Hyperband, fMRI and Silent imaging.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords