RF shimming for multi-transmit systems is commonly performed by phase-only shimming. Improved B1+ homogeneity is reached by phase-amplitude shimming but this requires knowledge of the B1+ magnitude distributions. For cardiac imaging at 7 Tesla, the acquisition of these distributions is challenging. We present a DREAM-based acquisition series to reconstruct B1+ magnitude maps in the heart. This method is applied to homogenize the transmit field for three subjects using phase-amplitude shimming. Results demonstrate a clear improvement of transmit field homogeneity in the heart in comparison to phase-only