A phase-rotation acquisition scheme is able to separate different signals in PRESS. In this work phase-rotation PRESS measurements were complemented by simulations to assess origins of spectral distortions in cardiac spectroscopy. Simulations show that, when using a phase-rotation acquisition scheme for PRESS without spoilers, more motion is acceptable when pulses are better calibrated. However, because of non-ideal pulse profiles, spoilers cannot be fully omitted if moving tissue is studied. It is shown that phase-rotation acqusition can, by combining simulations and measurements, be used as an elegant tool to investigate spectral distortions and optimize sequence parameters in PRESS of moving tissue.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.
Keywords