Abstract #5651

Controlling magnetisation transfer and exchange effects in mcDESPOT relaxometry.

Daniel J. West1, Rui Pedro A. G. Teixeira1, Jacques-Donald Tournier1, Joseph V. Hajnal1, and Shaihan J. Malik1

1Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom

Magnetisation transfer (MT) has been found to contribute to parameter estimation bias in sequences used in the two-pool relaxometry technique, mcDESPOT. Recent work shows that using controlled saturation magnetisation transfer (CSMT) RF-pulses, a two-pool system (free- and bound-pools) can behave as a single pool with modified equilibrium magnetisation and longitudinal relaxation rate. Here, we investigate the use of CSMT to model MT-effects in a three-pool system (exchange + MT) and show that under these conditions, signal behaves as a two-pool system and matches a mcDESPOT model but characterised by different parameters than those originally assumed.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords