Meeting Banner
Abstract #1437

Impact of Huygens Box in 3T MRI RF Safety Assessment

Xin Huang1, Xi Lin Chen2, and Shiloh Sison1

1Abbott, Sunnyvale, CA, United States, 2Abbott, Sylmar, CA, United States

Huygens Box is an efficient simulation technique to reduce simulation time and storage space. This paper uses simulations to investigate the impact of Huygens box on 3T MRI RF Safety Assessment. The numerical results on ASTM phantom shows the overall Symmetric Mean Absolute Percentage Error (SMAPE) average on typical MRI RF simulation is 6.28%.