Image distortion is a major limitation in radiation therapy (RT) planning, especially for diffusion weighed imaging in regions with strong B0-inhomogeneity. In this study we analyze the improvement of Readout-segmented-EPI over conventional single shot EPI in the geometrically challenging anatomical region of the neck. RS-EPI effectively increases geometric accuracy in head and neck tumor DWI and significantly reduces ghosting artifacts at the cost of a slightly prolonged acquisition time. Therefore it has proven a clear clinical benefit compared to standard SS-EPI.