We developed a novel MRI pulse sequence RAS-Q T1T2 for the simultaneous quantification of T1 and T2 of the myocardium using transient bSSFP imaging with a variable flip angle scheme. RAS-Q T1T2 was systematically analyzed based on a numerical simulation, as well as phantom and patient studies. We show that RAS-Q T1T2 yields accurate T2 estimates for the myocardium with a trade-off in precision compared to state-of-the-art methods, reducing scan time to less than 4s. The estimated T1 values reveal lower accuracy and precision than clinically established methods and need further improvements.