Meeting Banner
Abstract #2268

Damping of acoustic noise for neonatal MRI at 7 Tesla

Erik Huijing1, Evita Wiegers1, Dennis Klomp1, Fredy Visser1,2, Edwin Versteeg1, Koenraad Rhebergen3, Kim Annink4, Niek van der Aa4, Floris Groenendaal4, Jeroen Dudink4, Thomas Alderliesten4, Maarten Lequin1, Manon Benders4, and Jannie Wijnen1
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Philips HealthCare, Best, Netherlands, 3Otorhinolaryngology and Head & Neck Surgery, University Medical Center Utrecht, Utrecht, Netherlands, 4Neonatology, University Medical Center Utrecht, Utrecht, Netherlands

We designed a new acoustic hood for neonatal MRI exams at 7 Tesla. The new design is lightweight and allows easy access to the neonate during the MRI examination. We were able to optimize the new acoustic hood to a noise reduction of 10.6dB, exceeding the damping of a prototype with damping of 3.1dB. The new hood was designed to fit a head coil and allows the use of a ventilation tube.