Optimisation of multi-compartment relaxometry myelin water imaging (MCR-MWI)

Kwok-Shing Chan1 and José P. Marques1
1Donders Institute for Brain, Cognition and Behaviour, Radboud University, Nijmegen, Netherlands

In this study, we optimise the variable flip angle acquisition protocol for GRE-based myelin water imaging using MCR model. We show numerically and via in vivo imaging that MCR-MWI can be performed with as few as 3 flip angles and still be able to achieve similar measurement performance as with 7 flip angles. We also conduct an analysis to investigate the usefulness of an image denoising method to improve MWI. Preliminary results suggest that MCR-MWI is very robust to noise and higher resolution data is needed to have a better insight.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.