LCCC Balun for RF Coil

Xinqiang Yan1,2 and John C. Gore1,2
1Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 2Department of Radiology and Radilogical Science, Vanderbilt University Medical Center, Nashville, TN, United States

We introduced a novel LCCC lumped-element balun to suppress the unwanted common-mode current in RF coils and RF systems. Unlike the 4-element Lattice balun consisting of two capacitors and two inductors (LCCL), this design has 3 capacitors and only 1 inductor. With fewer inductors (previous lumped-element baluns need at least 2 inductors), this LCCC balun can further be miniaturized and has lower insertion loss.