Maximizing shim performance, particularly for larger organs such as the liver, is crucial at ultra-high field MRI due to the increased sensitivity to B0 field inhomogeneity. The conventional shimming in most MRI systems at ultra-high field, provides insufficient correction. However, by combining the available 2nd order spherical harmonic fields with an external array of 16 local shim coils, the magnetic field homogeneity in the liver is improved by as much as 44%.