Optimizing DWI b-value Sampling for Accurate Metabolic and Cytometric Parameter Extraction: Activity MRI [aMRI]

Xin Li1, Eric M. Baker1, Brendan Moloney1, Cory Wyatt1, Eric Baetscher1, Erin W. Gilbert2, Charles S. Springer1, Alexander R. Guimaraes1,3, and William D. Rooney1
1Advanced Imaging Research Center, Oregon Health & Science University, Portland, OR, United States, 2Surgery, Oregon Health & Science University, Portland, OR, United States, 3Diagnostic Radiology, Oregon Health & Science University, Portland, OR, United States

The optimal data acquisition strategy for accurate metabolic and cytometric parameter extraction using a digital DWI library is investigated using simulations under the constant data-acquisition time (same number of b-values) constraint. Base on DWI data from pancreatic tail tissue and using a seven b-value approach, the optimal maximum b-value is found to be data signal-to-noise ratio dependent, and generally should be below 3,000 s/mm2. In addition, evenly-spaced b-values are generally within the optimal b-spacing range.