Meeting Banner
Abstract #1814

Quantitative MRI measurements for T1 and T2 at 0.064T

Kalina V Jordanova1, Michele N Martin1, Karl F Stupic1, Samantha By2, Megan E Poorman2, Rafael O'Halloran2, and Kathryn E Keenan1
1NIST: National Institute of Standards and Technology, Boulder, CO, United States, 2Hyperfine, Inc., Guilford, CT, United States

Synopsis

Increased use of low-field MRI and quantitative MRI suggests a need for accurate quantitative measurements for fields <0.55T. We measure T1 and T2 for normal brain tissues at 0.064T using a Hyperfine scanner. T1 measurements gave a mean of 266±8.4ms for white matter, 358±6.6ms for grey matter; T2 measurements gave a mean of 95.8±2.4ms for combined white matter and grey matter, 794±357ms for CSF. Validation samples are measured on the Hyperfine system and compared to measurements taken on an electromagnet NMR at 0.064T.

This abstract and the presentation materials are available to members only; a login is required.

Join Here