Keywords: Motion Correction, Motion Correction
Motivation: PROPELLER/BLADE is robust to motion in brain imaging but comes at the cost of longer acquisition time.
Goal(s): Our goal was to reduce the acquisition time of commercial sequences (BLADE and EPI) based brain motion-insensitive workflow by a factor of 2.
Approach: The SMS-TGSE-BLADE sequence was developed with acceleration techniques, including in-plane GRAPPA, SMS, and EPI readout.
Results: Comparable image quality was obtained with the SMS-TGSE-BLADE sequence with more than 2-fold decrease in acquisition time.
Impact: The improvement in acquisition speed in the motion-insensitive brain examination (including T1-FLAIR, T2W, T2-FLAIR and DWI) through SMS-TGSE-BLADE sequence may increase patient comfort. It can also increase patient throughput and cost efficiency of healthcare providers.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.