Meeting Banner
Abstract #0485

Fast spin-echo triple-echo dixon (fTED) for fat suppression at 0.5T in the presence on knee implant.

Sajith Rajamani1, Harsh Kumar Agarwal1, Arjun Narula2, Uday Patil3, and Ramesh Venkatesan1
1GE HealthCare, Bangalore, India, 2Narula Diagnostics, Rohtak, India, 3Manipal Hospital, Bangalore, India

Synopsis

Keywords: Whole Joint, Low-Field MRI

Motivation: Fat saturation in musculoskeletal MRI at 0.5T in the presence of knee implants.

Goal(s): Echo spacing requirement of Fast spin-echo triple-echo Dixon (fTED) renders it as a preferred option for fat suppression even in the presence of implants without significant scan time penalty.

Approach: Commercially available fTED MRI was used to acquire 3 echoes (intra echo spacing of 6.71msec) between two consecutive refocusing pulses of FSE data acquisition at 0.5T research MRI system.

Results: Good quality fat suppression was observed in PD fat saturation knee MRI over 3 patients with knee implants.

Impact: Low and mid field MRI can benefit from fTED algorithm for fat suppression where water and fat are resonating in close proximity (in Hz) of each other without significant scan time penalty.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords