Keywords: Low-Field MRI, RF Arrays & Systems
Motivation: RF Coil design for low frequencies.
Goal(s): Demonstrate the impact of losses in different matching network topologies
Approach: we compared three- and four-element matching networks on 12 cm diameter coils at 10, 23.55, and 63.6 MHz. and evaluated coil loss and SNR.
Results: At 10 and 23.55MHz, the coils with three-element (single stage) matching networks exhibited approximately 50% lower unloaded Q compared to coils with four-element matching networks, whereas unloaded Q values were within 3% for three- and four-element matched coils at 63.6MHz. SNR at 23.55MHz (0.55T) was approximately 30% higher for the four-element matched coil over the three-element matched coil.
Impact: Resistive losses in four-element matching networks can be significantly lower than those of three-element networks at low frequencies or for lightly loaded coils, resulting in higher quality factor and SNR.
How to access this content:
For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.
After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.
After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.