Improved receive coil signal combination for large volume single voxel spectroscopy at ultra-high field

Mark A Elliott1, Neil Wilson1, Sophia Swago1, Ravi Prakash Nanga1, Ravinder Reddy1, and Walter Witschey1
1CAMIPM, Dept. of Radiology, University of Pennsylvania, Philadelphia, PA, United States

Synopsis

Keywords: Image Reconstruction, Spectroscopy, ultrahigh field

Motivation: Spectroscopic detection of low concentration metabolites (< 1mM) requires either lengthy signal averaging or large volume voxels. The latter can lead to significant variations in B0 across the volume, especially at ultrahigh field.

Goal(s): Correct for B0 heterogeneity in large volume SVS when detecting with multi-channel receive coils.

Approach: Using a metric for receive signal similarity, apply frequency and phase alignment of individual channel signals, prior to coil combination.

Results: The method was applied to 18 in vivo scans of 1H brain spectra at 7 T. Narrower linewidths and increased SNR were observed for water and down-field NAD+ resonances.

Impact: This approach provides an automated and robust method to improve spectral resolution and signal-to-noise ratio with large volume single voxel spectroscopy of low concentration metabolites when using multichannel receive arrays.

How to access this content:

For one year after publication, abstracts and videos are only open to registrants of this annual meeting. Registrants should use their existing login information. Non-registrant access can be purchased via the ISMRM E-Library.

After one year, current ISMRM & ISMRT members get free access to both the abstracts and videos. Non-members and non-registrants must purchase access via the ISMRM E-Library.

After two years, the meeting proceedings (abstracts) are opened to the public and require no login information. Videos remain behind password for access by members, registrants and E-Library customers.

Click here for more information on becoming a member.

Keywords