Keywords: MR-Guided Focused Ultrasound, RF Arrays & Systems, Focus Ultrasound
Motivation: MRI with the focused ultrasound treatment on the spine has low image quality, presumably due to a tailored receiver coil array.
Goal(s): A receiver coil array for 3T MRI is developed to image the spine of a large animal during focused ultrasound (FUS) treatment using a flat, large-aperture (25 cm diameter) phased array transducer.
Approach: A 4-channel receiver coil array was built with a housing curved to conform to the back of the specimen.
Results: Structural images with T1- and T2-weighted contrast confirmed an SNR gain of approximately 50% using this coil array compared with images acquired using a body coil.
Impact: The FUS-compatible spine MRI receiver coil array will enhance visualization of BSCB openings. As a clinical application, this coil array will also alleviate intraprocedural imaging challenges.